PL3140 series
datasheet

/ Short picosecond pulse mode-locked lasers
/ Up to 80 mJ
/ 10 ps pulses
/ 5-10 Hz pulse repetition rate
/ Short picosecond pulse mode-locked lasers
/ Up to 80 mJ
/ 10 ps pulses
/ 5-10 Hz pulse repetition rate
Features & Applications
Features
- Fiber master oscillator
- Diode pumped regenerative amplifier
- Flashlamp pumped power amplifier producing up to 80 mJ per pulse at 1053 nm
- 10 ps pulse duration
- Excellent pulse duration stability
- Up to 10 Hz repetition rate
- PC control via USB (RS232 is optional) and LabView™ drivers
- Remote control pad
- Optional streak camera triggering pulse with <10 ps rms jitter
- Optional thermostabilized second, third or fourth harmonic generators
- Optical parametric generators for tunable wavelength output in 210 – 2600 nm range are available
Applications
- Time resolved spectroscopy
- Nonlinear spectroscopy
- OPG pumping
- Other spectroscopic and nonlinear optics experiments
Description
Nd:YLF mode-locked PL3143 series picosecond lasers produces high energy pulses with as short as 10 ps pulse duration.
Rugged and reliable design
Diode pumped mode-locked quasi‑CW master oscillator produces the train of the pulses that is guided to the regenerative amplifier for further amplification. The single pulse is cavity‑dumped from regenerative amplifier and then amplified by linear amplifiers to up to 80 mJ pulse energy. The output pulse energy can be adjusted in approximately 1 % steps from 1 mJ to nominal output, at the same time pulse-to-pulse energy stability remains less than 1.5 % rms at 1053 nm.
Angle-tuned KD*P and KDP crystals mounted in thermostabilised ovens are used for second, third and fourth harmonic generation. Harmonics separators ensure high spectral purity of each harmonic directed to different output port. Build in energy monitors continuously monitors output pulse energy. Data from the energy monitor can be seen on the remote keypad or on PC monitor.
The laser provides triggering pulse for synchronization of customer‘s equipment with lead up to 500 ns. The lead of triggering pulse can be adjusted in ~0.25 ns steps from control pad or PC.
PRETRIG option is offered for streak camera triggering and can provide pulse with up to 1000 μs lead that can be adjusted from PC with approx. 33 ns step.
Simple and convenient laser control
For customer convenience the laser can be controlled via user-friendly remote control pad. The remote pad allows easy control of all the parameters and features a backlit display that is easy to read even when wearing laser safety eyewear. Alternatively, the laser can be controlled from personal computer via USB port using supplied software for Windows™ operating system. LabView™ drivers are supplied as well.
Specifications & Options
Model | PL3143 | PL3143A | PL3143B |
---|---|---|---|
Main specifications 1) | |||
Pulse energy | |||
at 1053 nm | 30 mJ | 50 mJ | 80 mJ |
at 526.5 nm 2) | 15 mJ | 25 mJ | 40 mJ |
at 351 nm 3) | 8 mJ | 12 mJ | 15 mJ |
at 263 nm 3) | 4 mJ | 6 mJ | 8 mJ |
Pulse energy stability (Std. Dev) 4) | |||
at 1053 nm | < 1.5 % | ||
at 526.5 nm | < 3.0 % | ||
at 351 nm | <5.5 % | ||
at 263 nm | <7.0 % | ||
Pulse duration (FWHM)5) | 10 ± 2 ps | ||
Pulse duration stability 6) | ± 0.5 ps | ||
Repetition rate | 10 Hz | 5 or 10 Hz | 5 Hz |
Polarization | linear, vertical | ||
Pre-pulse contrast | > 200:1 | ||
Triggering mode | internal/external | ||
SYNC OUT pulse jitter 7) | < 30 ps | ||
SYNC OUT pulse lead/delay 8) | -500..50 ns | ||
Beam divergence 9) | < 0.7 mrad | < 0.6 mrad | < 0.6 mrad |
Beam pointing stability 10) | <20 µrad | ||
Beam diameter 11) | ~6 mm | ~7 mm | ~8 mm |
Physical characteristics | |||
Laser head size (W × L × H) | 462 × 1245 × 255 mm | 600 × 1600 × 260 mm | |
Electric cabinet size (W × L × H) | 550 × 600 × 835 mm | ||
Umbilical length | 2.5 m | ||
Operating requirements | |||
Water consumption (max 20 °C) | < 15 l/min | ||
Room temperature | 22±2 °C | ||
Relative humidity | 20 – 80 % (non-condensing) | ||
Power requirements 12) | three phase, 208 or 380 VAC, 20 A, 50/60 Hz | ||
Power consumption | <2.5 kVA | <3 kVA13) | <4 kVA |
1) Due to continuous improvement, all specifications are subject to change without notice. Parameters marked typical are not specifications. They are indications of typical performance and will vary with each unit we manufacture. Unless stated otherwise all specifications are measured at 1053 nm.
2) For -SH option. Outputs are not simultaneous. Please inquire for pulse energies at other wavelengths.
3) With auxiliary H400 harmonics generator unit. Outputs are not simultaneous. Please inquire for pulse energies at other wavelengths.
4) Averaged from 300 pulses.
5) Inquire for optional pulse durations in 20 – 80 ps range.
6) Measured over 1 hour period when ambient temperature variation is less than ±1 °C.
7) In respect to optical pulse. <10 ps jitter is provided with PRETRIG option.
8) SYNC OUT lead or delay can be adjusted with ~0.25 ns steps in specified range. PRETRIG option provide -1000..5000 μs lead/delay time adjustment range.
9) Full angle measured at the 1/e² point at 1053 nm.
10) RMS value measured from 300 shots.
11) Beam diameter is measured at 1053 nm at the 1/e² point.
12) Mains voltage should be specified when ordering.
13) For 10 Hz version.
Options
PRETRIG option provides low jitter pulse for streak camera triggering with delay in -1000…5100 µs range and <10 ps rms jitter.
Performance & Drawings
Ordering information

PL3140 ordering information